What is OQM?
- OQM is an arrangement of software and sensors that collectively monitors the overall status of the instrument.
What does OQM do?
- Monitors the result of each calibration verification
- A calibration is performed if drift values exceed acceptable limits.
- Performs calibration and QC measurements to verify performance
- Initiates Corrective actions to correct any performance related issues.
- Disables any sensor that exhibits unexpected behavior
- Alerts users if any error occurs that will prevent operation
What is Calibration verification?
- Occurs every 30 minutes using the calibration solution stored in the cassette
- takes 30 seconds.
Does the OPTI R perform post measurement calibration verification?
- Occurs after each patient sample
- cassette is washed and filled with fresh calibration solution.
- takes 90 seconds.
How often does the OPTI R perform a Master Calibration?
- Occurs every 6 hours using the calibration solution and the gas bottle.
- takes 2 minutes to complete.
Does the OPTI R perform Auto-QC?
- The OPTI R contains 3 different levels of liquid QC material for use in quality control and quality monitoring.
- Auto-QC can be programmed to run on a custom schedule based on the needs of the customer.
Does the OPTI R need daily or weekly maintenance?
- The OPTI R is designed for ease-of-use and maximum uptime. This analyzer is virtually maintenance-free.
|